Vector network analyzers are both signal generators and receivers, so they have a large number of very necessary technical specifications. In this section, you will learn about some of the key technical specifications of network analyzers.
Maximum frequency
The maximum frequency of a VNA is the highest frequency it can measure. The receiver side of a network analyzer has an analog-to-digital converter (ADC) that converts the input signals into digital format. These signals can then be analyzed and displayed. However, the ADC does not have the ability to convert signals in the RF range, so the incident signal must be downconverted to its operating frequency. This operating frequency is called the intermediate frequency (IF).
dynamic range
Dynamic range is the range of power over which the response of a component can be measured.
This figure shows two different ways of defining dynamic range. The system dynamic range is the value used for instrument specifications.
The system dynamic range is the function of the instrument when no boost amplifier is used and the gain of the device under test is not taken into account. The maximum source power of an instrument is its maximum power level, Pref
The receiver dynamic range is the dynamic range of the instrument when power amplification is used. Unlike using the source power as the maximum power level, this specification is based on the maximum power Pmax that can be measured at the receiving end of the instrument.
Defining the dynamic range
A trace measured by the bandpass filter S21, which shows the dynamic range of the instrument, is shown in the lower left-hand plot. The upper limit of the trace is relatively flat and the lower limit contains some noise. Let's look at what factors determine the shape of these boundaries.
The maximum power level of the dynamic range is determined by the upper limit of the source power level and the compression point of the receiver.
The mixers and amplifiers that make up a receiver can only handle so much power before they reach saturation, or before they reach maximum output. When a device is in the saturation region, there is no longer a linear relationship between its input and output.
The saturation of the amplifier can be seen in the right-hand diagram below. At input powers higher than 1W, the actual output (red) will deviate from the desired output (green). This phenomenon is called compression. The receiver cannot capture the signal from any device output above its compression point. This limitation of the input power constitutes the upper limit of the dynamic range.
Dynamic range of traces In the gain compression plot, the ideal linear transfer function of the amplifier is shown in green and the true transfer function is shown in red.
output power
Output power reflects how much power the VNA's signal generator and tester can transmit into the device under test. It is expressed in dBm and is referenced to 50Ω impedance to match the characteristic impedance of most RF transmission lines.
High output power is useful for improving the signal-to-noise ratio of a measurement or determining the compression limits of the device under test.
Many active devices, such as amplifiers, require extremely challenging linear and nonlinear high power measurements beyond the power limits of the network analyzer.
trace noise
Trace noise is the superimposed noise you see forming on the response of the device under test due to random noise in the system. It can make the signal look less smooth and even a little jittery.
Trace noise can be eliminated by increasing the test power, decreasing the bandwidth of the receiver, or averaging.
Many people use digital multimeters to easily measure current, voltage, resistance, etc. However, there is a drawback to using a digital multimeter to measure current, namely: the circuit must be disconnected and the digital multimeter must be connected in series with the circuit under test in order to make the measurement. In this case, a current probe can be used instead of a digital multimeter to make measurements. Since the electrical connection of the circuit is no longer required, the current probe is more time-consuming...
A multimeter is an instrument that can measure a wide range of parameters, such as alternating (AC) or direct current (DC) voltage, AC or DC current, resistance, temperature, capacitance, and so on. Modern digital multimeters use digital and logic technology to become smaller and smaller with more and more built-in functions. By introducing digital technology, digital multimeters can provide graphical displays, data logging and signal digitization as well as...
A programmable DC power supply is connected to the device under test through wires connected to a DC power panel. Engineers can set voltage or current levels to power the device for testing. The DC power supply is a programmable DC power supply.
Automated Test Equipment (ATE) system power supplies are programmable power supplies designed specifically for ATE systems. Form factor is critical. ATE power supplies with higher power densities take up less rack space, reducing the factory floor space required for test equipment. The ATE power supply is a programmable power supply designed specifically for ATE systems.