technical specification
frequency range | EM5080L | 9kHz to 30MHz |
power level (elec.) | ||
Maximum RF Level (CW) | RF attenuation ≥ 20 dB, RF preamplifier off. | 30dBm (=1W) |
Maximum pulse voltage | RF attenuation ≥20dB, RF preamplifier off | 150V |
Resolution bandwidth | ||
Analyzer Mode | 10Hz to 1MHz (-3dB) using 1/2/3/5/10 steps | |
receiver mode | Standard: 200Hz, 9kHz, 120kHz (-6dB)
Also added 10kHz, 100kHz (-6dB) selectable. |
|
preselector | Can be turned off in the analyzer | 15-way fixed filter (EM5080B) |
preamplifier | Can be turned on/off | 9kHz to 1GHz, 20dB gain |
measuring time | receiver mode | 1ms to 1.5s |
oscillograph | receiver mode | Peak, quasi-peak, average, |
Display Average Noise Level (DANL) | Receiver Mode, Normalized, Average Detector (AV), RF Attenuation 0dB , Typ. | |
Preamplifier off | ||
9kHz<f<150kHz, bandwidth 200Hz | <0dBμV | |
150kHz<f<30MHz, bandwidth 9kHz | <10dBμV | |
30MHz<f<1GHz, 120kHz bandwidth | <20dBμV | |
Preamplifier on | ||
9kHz<f<150kHz, bandwidth 200Hz | <-5dBμV | |
150kHz<f<30MHz, bandwidth 9kHz | <0dBμV | |
30MHz<f<1GHz, 120kHz bandwidth | <5dBμV | |
Total measurement uncertainty | 9kHz≤f≤1GHz | 1.5dB |
Power Consumption | <100W |