■ Simple operability

■ 6 test items or combinations, 9 test storage areas

■ Contact check function (TH9302C, TH9302D)

■ RS-232C interface as standard

■ Relay input/output HANDLER interface for PLC support

■ External control interface to support synchronized or remote control

■ Voltage rise time: 0.1s-999.9s

■ Test wait time: 0.1s-999.9s

Categories. , Tag. Brand.

Inquire about this product

TH9302 series AC/DC Withstand Voltage and Insulation Tester is a newly developed instrument with high cost-effective and simple operation by Tonghui Electronics.

The principle structure of the instrument: the current module is a DA reference, controllable sinusoidal generator, class AB amplifier, output current closed-loop control DA reference: to ensure that the amplitude of the output voltage can be controlled controllable sinusoidal generator: in the AC output can be set to work in the 50Hz, 60Hz, no longer subject to the limitations of the line voltage Line Power Amplifier: the voltage waveform of the distortion of a small degree of simplicity, simple control, high reliability of the anti-impact capability

All series are equipped with HANDLER, RS-232C interface, to adapt to a variety of different needs of high security and reliability of the automatic test system

Summary parameters TH9302 TH9302C TH9302B TH9302D
pressure resistance alternating current 0.10kV-5.00kV
alternating current (electricity) 0.01mA-12mA
dc voltage 0.10kV-6.00kV ------
direct current 0.01mA-5mA ------
isolate input voltage 0.10V-1.00kV ------
resistive 0.02MΩ-10GΩ ------
Rear Panel Output ------ there are ------ there are
exposure check ------ there are ------ there are

A. Rear output holes to support contact checks (CK)

For the high-intensity test, due to the internal breakage of the test lead, poor contact of the test tooling, etc. caused by the test withstand voltage, insulation misjudgment failed, TH9302C/TH9302D two models in the rear panel of the instrument to increase the high-voltage test end with the CK function, to ensure that the instrument test results of the validity and correctness of the instrument can be avoided due to the loss of the test accessory and tooling to the customer brought about by the misjudgement.

        

         A description with words on the screen has been automatically generated

B. Optocoupler isolated relay output HANDLER, directly connected to the PLC, without conversion circuitry

 

Product Model TH9302 TH9302C TH9302B TH9302D
Pressure resistance test
voltage output AC voltage range 0.10kV-5.00kV
Voltage waveform simple harmonic vibration
distortion <3%
operating frequency 50, 60Hz selectable
Frequency accuracy ±1%
output power 60VA (5.00kV12mA)
Voltage Adjustment Ratio ± (2.0%+50V) (rated power)
DC voltage range 0.10kV-6.00kV ------
signal source frequency 600Hz ------
output power 25VA (5.00kV5mA) ------
Voltage Adjustment Ratio ± (2.0%+100V) (rated power) ------
Voltage Resolution 10V
Voltage Accuracy ±(2.0% setting +20V)
Voltage Generation Method DDS signal source plus Class AB amplifier
current test AC current range 0.01mA-10mA
Short-circuit current (instantaneous) >20mA (set output voltage >500V)
Current Resolution 0.01mA
Current Accuracy ± (2% reading + 2 words)
Actual current OFF-0.01mA-10mA
DC current range 1mA-5.00mA ------
Current Accuracy ± (2.01 TP3T reading + 2 words) ------
Discharge function (DCW automatically discharged at the end of the test) ------
Insulation resistance test
output voltage 0.10V-1.00kV ------
Voltage Resolution 10V ------
Voltage Accuracy ± (2.0% reading + 10V) ------
Maximum Output Current 5mA ------
Output instantaneous short-circuit current >10mA (set output voltage >500V) ------
Load Adjustment Ratio ≤1% (rated power) ------
Resistance Measurement Range 0.02MΩ-10GΩ ------
Resistance display range

(1000V)

5mA 0.2MΩ-1MΩ ------
1mA 1MΩ-10MΩ ------
100uA 10MΩ-100MΩ ------
10uA 100MΩ-1GΩ ------
1uA 1GΩ-10GΩ ------
Resistance Measurement Accuracy ≥500V

1MΩ-1GΩ ± (5% reading + 5 words)

1GΩ-10GΩ ± (10% reading + 5 words)

<500V

0.1MΩ-1GΩ ± (10% reading + 5 words)

1GΩ-10GΩ is for reference only with no accuracy requirement.

------
Current measurement accuracy ± (2% reading + 2 words) ------
arc detection
Measurement range AC 1mA-20mA
DC 1mA-20mA ------
comparator
discriminant Window Comparison Method (Condition I)arrive at (a decision, conclusion etc)<Ifirst (of multiple parts))
Iarrive at (a decision, conclusion etc)ON: When Iarrive at (a decision, conclusion etc)<Ix<Ifirst (of multiple parts)PASS;
When Ix≤Iarrive at (a decision, conclusion etc)or Ix ≥ Ifirst (of multiple parts)FAIL
Iarrive at (a decision, conclusion etc)OFF: When Ix<Ifirst (of multiple parts)PASS;
When Ix ≥ Ifirst (of multiple parts)FAIL
Insulation resistance is determined in the same way as above
Current limit setting Ifirst (of multiple parts) AC 0.01mA-10mA
DC 1uA-5mA ------
Lower current limit setting Iarrive at (a decision, conclusion etc)

(LOWEROFF)

AC 0.01mA-10mA
DC 1uA-5mA ------
Resistance limit setting OFF, 1MΩ-10GΩ ------
Lower Resistance Limit Setting 0.1MΩ-10GΩ ------
discriminant output PASS/FAILLCD and LED display, audible alarm.
Other Functions
Rear Panel Output Hole ------ there are ------ there are
exposure check ------ be in favor of ------ be in favor of
parameterization
Voltage Rise Time 0.1s-999.9s
Test Time Setting 0.3s-999.9s (at TIMERON)
time accuracy ± (0.2% setpoint ± 0.1s)
measurement function
keypad lock Prevent accidental modification of test conditions, or prohibit test conditions from being modified
Current over range judgment Hardware quickly determine insulation collapse, faster and safer than voltage sampling, reducing impact damage to the product
ARC Arc Detection Sampling of current burst signals to determine the potential hazards and approximate size of the loop
High Pressure Indication Window indication and LED light indication
Storage and Interfaces
File programming and storage Programmable 9 test files
USB port not have
Power-up parameter saving Setting parameters are saved as default parameters and can be restored automatically next time the power is turned on.
control interface HANDLER
communication interface RS232C, RS485
General technical indicators
Operating temperature, humidity 0℃-40℃,≤80%RH
power supply 100V-121V, 198V-242V47.5-63Hz
power wastage ≤300VA
external volume 280mm×89mm×402mm
weights Approx. 12kg

■ Simple operability

■ 6 test items or combinations, 9 test storage areas

■ Contact check function (TH9302C, TH9302D)

■ RS-232C interface as standard

■ Relay input/output HANDLER interface for PLC support

■ External control interface to support synchronized or remote control

■ Voltage rise time: 0.1s-999.9s

■ Test wait time: 0.1s-999.9s

■ Automated test systems

▪ Household appliances

■ Transformers, motors

■ Electrical equipment

▪ Electric heating appliances

▪ Lighting industry

▪ New energy vehicles

■ Electronic components

▪ Medical equipment

Expand More