Electrical Characteristics
Measurement conditions:23℃,60%RH,No current carrying wire nearby, tested with the wire under test through the center, load impedance 1MΩ.
| gear level (i.e. first gear, high gear etc) | L | H | ||
| current range | 50mA~10A peak | 1A~100A peak | ||
| Range Sensitivity | 0.1V/A | 0.01V/A | ||
| Typical DC Accuracy | 3%±50mA | 500mA~40A peak: 4%±50mA; 40A~100A peak: ±15% max. | ||
| bandwidths
(-3dB) |
CPL8100A | DC-600kHz Reference Typical Amplitude-Frequency Characteristics (Figure 1) | ||
| CPL8100B | DC-2MHz reference typical amplitude-frequency characteristics (Figure 2) | |||
| Maximum current vs. frequency characteristic curve | Refer to Figure 3 | |||
| phase shift | DC~65Hz: <1.5° | DC~65Hz: <1°. | ||
| Typical DC Linearity | Typical DC linearity at H-stop (0.01V/A) Refer to Figure 4 | |||
| rising time | CPL8100A | ≤583ns | ||
| CPL8100B | ≤175ns | |||
| Maximum operating current | See table of voltage and current ratings | |||
| Maximum working voltage | See table of voltage and current ratings | |||
| Maximum Floating Voltage | See table of voltage and current ratings | |||
| Operating voltage RMS | cati 600v catii 600v catiii 300v | |||
| Common mode voltage RMS | cati 600v catii 600v catiii 300v | |||
| Typical battery type and life | 9V Alkaline Laminated Battery / 15 hours | |||
| Low battery indicator function | When the battery voltage <6.5V, the battery indicator is red. | |||
| Overload indication function | Measured current exceeds range, buzzer sounds | |||
Comprehensive test and measurement service provider-Shenzhen Weike Electronic Technology Co.