Product Overview: 725Ex Intrinsically Safe Multifunction Process Calibrator
Simple and powerful intrinsically safe calibration tool
The new 725Ex intrinsically safe multifunction process calibrator is powerful and easy to use. Coupled with one of the new Fluke-700PEx pressure modules (8), the 725Ex is capable of calibrating virtually any maintenance-required process equipment used in areas where explosive atmospheres are present.
Measure and source milliamps, voltages, temperatures (RTD and thermocouple), frequencies, ohms, and pressures using the Pressure Module option.
Do you need to calibrate a transmitter?The 725Ex split-screen display lets you view input and output values at the same time. For valve and I/P testing, you can source milliamps while measuring pressure.The 725Ex features auto-stepping and auto-sloping for remote testing, and 25% stepping for fast linear testing.
The Fluke 725Ex is ready to take and use. The controls are simple, menu-free and easy to operate. Its memory function enables quick setup. Its rugged design allows you to use it as you wish.
Product Specifications: 725Ex Intrinsically Safe Multifunction Process Calibrator
| measurement accuracy | ||
| dc voltage | 30.000 V | 0.02% + 2 words (top of display) |
| 10.000 V | 0.02% + 2 words (lower display) | |
| 90.00 mV | 0.02% + 2 characters | |
| -10.00 mV to 75.00 mV | 0.025% + 1 count (via TC connector) | |
| direct current | 24.000 mA | 0.02% + 2 characters |
| resistive | 0.0 to 400.0 Ω | 0.1 Ω (4-wire), 0.15 Ω (2- and 3-wire) |
| 401 to 1500 Ω | 0.5 Ω (4-wire), 1 Ω (2- and 3-wire) | |
| 1500 to 3200 Ω | 1 Ω (4-wire), 1.5 Ω (2- and 3-wire) | |
| frequency | 2.0 to 1000.0 CPM | 0.05% +1 character |
| 1.0 to 1100.0 Hz | 0.05% +1 character | |
| 1.00 to 10.00 kHz | 0.05% +1 character | |
| (level of) sensitivity | 1 V Minimum Peak to Peak | |
| stresses | Accuracy is 0.025% of range when using any of the 8 intrinsically safe pressure modules (see Pressure Modules in Options and Accessories for detailed specifications). The following modules are available: differential pressure, manometer, vacuum, absolute pressure, dual pressure, and high pressure. | |
| Output Accuracy | ||
| dc voltage | 100.00 mV | 0.02% + 2 characters |
| 10.000 V | 0.02% + 2 characters | |
| -10.00 mV to 75.00 mV | 0.025% + 1 count (via TC connector) | |
| direct current | 24.000 mA (output) | 0.02% + 2 characters |
| 24.000 mA (analog) | 0.02% + 2 characters | |
| resistive | 15.0 to 400.0 Ω | 0.15 Ω (excitation current 0.15 to 0.5 mA), 0.1 Ω (excitation current 0.5 to 2 mA) |
| 401 to 1500 Ω | 0.5 Ω (excitation current 0.05 to 0.8 mA) | |
| 1500 to 3200 Ω | 1 Ω (excitation current 0.05 to 0.4 mA) | |
| frequency | 2.0 to 1000.0 CPM | 0.05% |
| 1.0 to 1100.0 Hz | 0.05% | |
| 1.00 to 10.00 kHz | 0.25% | |
| waveform | 5 V p-p square wave, -0.1 V offset | |
| RTD and thermocouple | ||
| Measurement accuracy | NI-120 | 0.2 °C |
| PT-100 (385) | 0.33 °C | |
| PT-100 (393) | 0.3 °C | |
| PT-100 (JIS) | 0.3 °C | |
| PT-200 (385) | 0.2 °C | |
| PT-500 (385) | 0.3 °C | |
| PT-1000 (385) | 0.2 °C | |
| resolution (of a photo) | 0.1 °C | |
| J | 0.7 °C | |
| K | 0.8 °C | |
| T | 0.8 °C | |
| E | 0.7 °C | |
| R | 1.8 °C | |
| S | 1.5 °C | |
| B | 1.4 °C | |
| L | 0.7 °C | |
| U | 0.75 °C | |
| N | 0.9 °C | |
| resolution (of a photo) | J, K, T, E, L, N, U: 0.1 °C, 0.1 °F B, R, S: 1 °C, 1 °F | |
| XK | 0.6 °C | |
| BP | 1.2 °C | |
| Output Accuracy | NI-120 | 0.2 °C |
| PT-100 (385) | 0.33 °C | |
| PT-100 (393) | 0.3 °C | |
| PT-100 (JIS) | 0.3 °C | |
| PT-200 (385) | 0.2 °C | |
| PT-500 (385) | 0.3 °C | |
| PT-1000 (385) | 0.2 °C | |
| resolution (of a photo) | 0.1 °C | |
| Accuracy is suitable for 4-wire measurements. | ||
| J | 0.7 °C | |
| K | 0.8 °C | |
| T | 0.8 °C | |
| E | 0.7 °C | |
| R | 1.4 °C | |
| S | 1.5 °C | |
| B | 1.4 °C | |
| L | 0.7 °C | |
| U | 0.75 °C | |
| N | 0.9 °C | |
| resolution (of a photo) | J, K, T, E, L, N, U: 0.1 °C, B, R, S: 1 °C | |
| XK | 0.6 °C | |
| BP | 1.2 °C | |
| technical specification | ||
| Slope Function | output function | Voltage, current, resistance, frequency, temperature |
| slope | Slow Linear Slope, Fast Linear Slope, 25% Stepper Slope | |
| Loop Power Function | input voltage | 12 V |
| accuracy | 10% | |
| Maximum current | 22 mA, short circuit protected | |
| stepper function | output function | Voltage, current, resistance, frequency, temperature |
| stepper | Range of 25%, Range of 100% | |
| Safety specifications | ||
| Institutional approval | I.S. Class I, Division 1, Group B-D | |
| Mechanical and general specifications | ||
| sizes | 130 x 236 x 61 mm (5.188 x 9.291 x 2.402 in) | |
| weights | 0.85 kg (1.874 lbs.) | |
| batteries | 4 AA alkaline batteries | |
| warranty period | third year | |
| Battery replacement | Separate battery compartment for operation without breaking calibration seals | |
| Side Port Connections | Pressure Module Connectors | |
Model: 725Ex Intrinsically Safe Multifunction Process Calibrator
725Ex Intrinsically Safe Multifunction Process Calibrator
725Ex Intrinsically Safe Multifunction Process Calibrator
Versatile process calibrator includes:
- Set of AC72 alligator clips
- Set of retractable test leads
- Set of TL 75 test leads
- user manual
- NIST-traceable calibration reports and data
Comprehensive test and measurement service provider-Shenzhen Weike Electronic Technology Co.