DP-25 Active Differential Probe (1400V, 25MHz)
| model number | DP-25 |
| Bandwidth (-3dB) | DC-25 MHz (x50 steps, x200 steps)
DC-15MHz (x20 steps) |
| Attenuation ratio | x20,x50,x200 |
| rising time | 14ns (x50 gears, x200 gears)
23.4ns (x20 gears) |
| accurate | ±2% (after 20 minutes of warm-up) |
| Input Voltage Range | ≤140V at x20, ≤350V at x50, ≤1400V at x200 |
| Input DC Voltage | 700V |
| RMS | 500Vrms |
| mixed messages | ≤2.0mV rms |
| Input Impedance | Between the two ends: 4MΩ//1.2pF
Single-ended to ground: 2MΩ//2.4pF |
| Clutter suppression ratio (CMRR) | 60Hz:>80dB; 100Hz:>60dB; 1MHz:>50dB |
| operating temperature | 0~50℃ |
| Operating humidity | 10~85%RH |
| weights | 250g |
| sizes | 195*55*30 mm |

1400V:1KHz / 700V:10KHz / 300V:100KHz / 200V:1MHz / 150V:10MHz / 100V:25MHz


Comprehensive test and measurement service provider-Shenzhen Weike Electronic Technology Co.