TH2851 series impedance analyzer is a new generation of impedance test instrument successfully developed by Changzhou Tonghui Electronics using the current international advanced automatic balancing bridge principle, which is the latest height of domestic impedance test instrument.
TH2851 series impedance analyzer completely subverted the complex and cumbersome operation interface of traditional domestic instruments, based on Windows 10 operating system, realizing the full computerized operation interface, making the test smarter and simpler.
TH2851 series impedance analyzer also completely surpassed the 120MHz frequency bottleneck of similar foreign instruments; solved the defects of similar foreign instruments that can only be analyzed and cannot be individually tested; the Chinese and English operation interface also solved the embarrassment of the English interface of the foreign instruments only; adopted the two interfaces of single measurement and analysis, which makes the test simpler.
The fast 2.5ms test speed and impedance range of up to 40MΩ meets the measurement requirements of components and materials, and is especially beneficial for low loss (D) capacitors and high quality factor (Q) inductors. The four-terminal port configuration effectively eliminates the influence of electromagnetic coupling of the test line and extends the lower limit of low-impedance testing capability downward by a factor of ten compared to conventional five-terminal configurations.
The TH2851 series impedance analyzers include the following types:
| Summary parameters | TH2851-015 | TH2851-030 | TH2851-050 | TH2851-080 | TH2851-130 | |
| Test Frequency | 10Hz-15MHz | 10Hz-30MHz | 10Hz-50MHz | 10Hz-80MHz | 10Hz-130MHz | |
| Basic precision | 0.08% | |||||
| ACsignal source | input voltage | 5mVrms - 2Vrms | ||||
| amps | 50μA - 20mArms | |||||
| DCbias voltage | input voltage | 0V - ±40V | ||||
| amps | 0mA-±100mA | |||||
- highly accurate
The application of broadband automatic zeroing type auto-balancing bridge technology can achieve ideal 10% measurement accuracy in 10Hz-130MHz frequency, 25mΩ-40MΩ impedance range, of which the highest accuracy reaches 0.08%, which is much higher than the accuracy of impedance analyzers and network analyzers of RF reflection measurement method.
- High stability and consistency
The following figure shows the curve of measuring 100Ω resistance at speed 5 and test frequency 1MHz, and the trace noise ≦0.003% (≦±0.0015Ω) can be seen from the following figure
- high speed
- Large 10.1" screen with four measurement parameters for details at a glance
10.1-inch touch screen, 1280*800 resolution, Windows 10 system, Chinese and English operation interface, support for keyboard, mouse, LAN, VGA/HDMI interfaces, brings unparalleled ease of operation.
The large screen brings more benefits is that you can place all the test parameters and sorting parameters, sorting results, function selection and other parameters in the same screen, and it never looks crowded and cluttered, and at the same time, you can display four kinds of measurement parameters, four kinds of measurement parameters can be adjusted arbitrarily. The button on the left side of the screen can quickly select eight sets of test parameters
- Enhanced list scanning functionality
A list scan of up to 1601 points can be set, and test conditions such as test frequency, test voltage, DC bias, etc. can be set individually for each point.
- Powerful graphical interface for analysis
Up to 4 channels can be displayed simultaneously, and each channel can display up to 4 curves. There are fourteen display methods for each channel and curve.
| conduit | Display Window | Individual windows can be set up with different scanning conditions
4 curves can be set per window |
| in a roundabout way | measured parameter |
- Segmented Scanning Function
Segment scanning is to set different frequency segments for scanning in one scanning cycle, different levels and offsets can be set during scanning, and the scanning result is directly displayed graphically, which is used for scanning needs that require fast screening of multiple frequency band parameters.
If a crystal resonator needs to be tested for nominal resonance/anti-resonance frequencies as well as other spurious frequencies, the Segmented Sweep function allows the measurement to be scanned over a specific frequency range without scanning for uncorrelated frequencies.
- Powerful cursor analysis
TH2851 series precision impedance analyzer has a powerful cursor analysis capability, which can realize the following functions by cursor:
- Reading the value of the measurement result (as an absolute value or as a relative value with respect to a reference point)
- Finding a specific point on a curve (cursor find)
- Analyzing curve measurements and calculating statistics
- Use cursor values to modify the scanning range and vertical scale.
The TH2851 can display 10 cursors on each curve, including the reference cursor.
Each cursor has an excitation value (the value corresponding to the X-axis of the coordinate system) and a response value (the value corresponding to the Y-axis of the coordinate system).
The Cursor Find function allows searching for the following conditional measurement points:
| Maximum value, minimum value | |
| Peak and valley values: | Peaks (extreme values), valleys (very small values), nearest peaks and valleys to the left of the cursor, nearest peaks and valleys to the right of the cursor, multiple peaks and valleys |
| Target value: | Nearest target value to the cursor, nearest target value to the left of the cursor, nearest target value to the right of the cursor, multiple target values |
- Powerful graphical analysis
1) Curve sorting function
All or part of the scanning curve test value can be qualified / unqualified judgment, commonly used in resonance curve screening such as piezoelectric components and other resonance frequency.
2) Equivalent circuit analysis test
Different types of real-life devices can be equated to simple three-parameter, four-model, four-parameter, three-model impedance devices. The Equivalent Circuit Analysis test provides seven basic circuit models for equating these devices.
You can compare the impedance fitting curves of the simulated equivalent circuit parameter values with the actual measured impedance curves, and you can also fit them according to the selected model by the parameters you enter.
Equivalent circuit models can be directly exported to TXT files for users to save and use.
3) Crystal Oscillator Analysis (Piezo Device Analysis)
Measurement and performance analysis of a crystal oscillator is carried out to obtain important parameters such as resonant frequency, anti-resonant frequency, quality factor, etc. of the crystal after measurement and calculation.
At the same time, for other piezoelectric devices such as piezoelectric ceramic filters, piezoelectric ceramic traps, piezoelectric ceramic frequency discriminator, piezoelectric ceramic transformer, high-power ultrasonic generator, transducer (oscillator), surface acoustic wave devices, electroacoustic devices and so on can be tested for such as static capacitance, loss, resonance frequency, anti-resonance frequency, mechanical coupling coefficient and other parameters.
4) Comparison of curve trajectories
Curve Trajectory Comparison is used for continuous measurements of the measured part, with all curves displayed in the same coordinate system. By the following two applications:
- a) For a wide range of different DUTs
Comparison of curve trajectories under different measurement conditions
Set different frequency points and calculate the measured values of all curves at that frequency point
- b) For the same DUT
Repeatability of multiple measurements compared to measurements taken under the same conditions
Setting different frequency points and calculating the measured values of all curves at that frequency point
- Dielectric constant analysis (optional)
TH2851 series precision impedance analyzer support dielectric material dielectric constant, dielectric loss analysis, need to purchase test fixture and analysis software.
- configure
| serial number | model number | name (of a thing) | mandatory option |
| 1 | TH2851_*** | Impedance Analyzer Mainframe | R |
| 2 | TH26077 | Dielectric Material Test Fixture | R |
| 3 | TH2851 | Dielectric Constant Test Software | R |
- TH26077 Dielectric Material Test Fixture
The TH26077 Dielectric Material Test Fixture is designed to make accurate dielectric constant and dielectric loss measurements on the TH2851. Discrete capacitance is eliminated by measuring the dielectric constant of solid materials using the parallel plate method.
1) External structure
2) Basic parameters
| Test Frequency | DC-30MHz | ||
| measured parameter | Capacitance (C), depletion factor (D), dielectric constant (εr', εr'') | ||
| Test Electrodes | Electrode type | Diameter of measured material | Thickness of measured material |
| Shield Electrode A1 | 40mm - 56mm | ≤10mm | |
| Shielded electrode A2 | 10mm - 56mm | ≤10mm | |
| Test Principle | Parallel Plate Measurement | ||
| zero (a hard drive) | Open circuit, short circuit | ||
3) Test Principle
4) Dielectric constant formula
- Dielectric Constant Analysis Software
The software is optional, after opening the TH2851 interface directly select the corresponding parameters can be, respectively, single measurement, list scanning, curve scanning, the following figure is the curve scanning interface.
- Magnetic permeability analysis
TH2851 supports magnetic permeability testing of magnetic materials. On the basis of the mainframe, it needs to be equipped with optional test fixture and magnetic permeability analysis software, the specific configurations are as follows:
- configure
| serial number | model number | name (of a thing) | mandatory option |
| 1 | TH2851 | Impedance Analyzer Mainframe | R |
| 2 | TH26086 | Test Fixture Base | R |
| 3 | TH26088 | Magnetic Permeability Test Fixture | R |
| 4 | TH2851 Magnetic Permeability Test Software | R | |
| 5 | TH26086-CAL | Fixture Calibration Kit | R |
- Test Fixture
1) TH26086 adapter base
Used to convert a 4-end pair into an N-type 7mm base, adapted to other test fixtures such as the TH26088 Magnetic Permeability Test Fixture.
The TH26086 is a mandatory option, if not selected, the TH26088 will not be measured correctly.
Technical Parameters
| Test Frequency | DC - 130MHz |
| Maximum voltage | ±42V peak (AC+DC) |
| operating temperature | 0°C - 40°C |
2) TH26088 Magnetic Permeability Test Fixture
The TH26088 accurately measures the permeability of toroidal magnetic materials. Since the fixture is constructed with a toroidal coil wrapped around one turn (no magnetic leakage), there is no need for a wire wrapped around a toroidal coil, and it allows for a wide range of frequency coverage. It comes with one small and one large fixture to accommodate a wide range of sizes, making it an ideal structure for measuring magnetic materials.
Technical Parameters
| Test Frequency | 1kHz - 1GHz | ||
| Sample size | Height h | Inner diameter rangeb (diameter) | Outer diameter rangec (diameter) |
| For small parts | ≤ 3 mm | ≥ 3.1 mm | ≤ 8 mm |
| For large parts | ≤ 8.5 mm | ≥ 5 mm | ≤ 20 mm |
- Test Principle
1) Basic Concept of Magnetic Permeability
2) Effective permeability
Measured by the inductance measurement method, the wire is wound on the core of a toroidal core and placed in a closed cylindrical space (TH26088 fixture), the inductance (self-inductance) at both ends of the wire is measured by an impedance analyzer, and the effective permeability is deduced based on the test inductance and the following formula.
3) Relative permeability
- Magnetic permeability analysis software (optional)
After choosing to open the permeability analysis software, the instrument can directly select the permeability parameters in the single measurement, list scan and curve scan, and directly display the test results in the interface.
- Standard Accessories
- Optional accessories
| Product Model | TH2851-015 | TH2851-030 | TH2851-050 | TH2851-080 | TH2851-130 | |
| demonstrate | 10.1-inch TFT LCD display 1280 x RGB x 800, capacitive touchscreen | |||||
| AC Measurement Parameters | Cp/Cs, Lp/Ls, Rp/Rs, complex Z, |Z|, complex Y, |Y|, R, X, G, B, θ, D, Q, VACIAC | |||||
| DC measurement parameters | VDC, IDC, DCR | |||||
| Test Frequency | realm | 10Hz - 15MHz | 10Hz - 30MHz | 10Hz - 50MHz | 10Hz - 80MHz | 10Hz - 130MHz |
| highest resolution | 1mHz | |||||
| Frequency relative error | ≤±0.0007% | |||||
| test level | AC Voltage | 5mV - 2Vrms | ||||
| resolution (of a photo) | 1mV | |||||
| AC current | 50μA - 20mArms | |||||
| resolution (of a photo) | 10μA | |||||
| DC Bias | input voltage | 0V - ±40V | ||||
| Voltage Resolution | 1mV | |||||
| amps | 0mA - ±100mA | |||||
| Current Resolution | 40μA | |||||
| Test Side Configuration | four-pronged attack (e.g. marriage) | |||||
| Output Impedance | 25Ω/100Ω | |||||
| Typical measurement time (speed) | Five speeds: 1 (fast) - 5 (precise)
1:2.5ms 2:10ms 3:40ms 4:80ms 5:400ms (arithmetic average excluding communication time, test speeds will vary slightly for each frequency) |
|||||
| Maximum accuracy | 1kHz: 0.08%
1MHz: 0.08% 2MHz: 0.5% 10MHz: 1% 130MHz: 5.0% |
|||||
| Measurement display range | a 1 x 10-18E 1 x 1018 | |||||
| Cs, Cp | ±1.00000 aF - 999.999 EF | |||||
| Ls, Lp | ±1.00000 aH - 999.999 EH | |||||
| D | ±0.00001 - 9.99999 | |||||
| Q | ±0.01 - 9999.99 | |||||
| R, Rs, Rp, X, Z | ±1.00000 aΩ - 999.999 EΩ | |||||
| G, B, Y | ±1. 00000 aS - 999.999 ES | |||||
| Vdc | ±1.00000 aV - 999.999 EV | |||||
| Idc | ±1.00000 aA - 999.999 EA | |||||
| θr | ±1.00000 a rad - 3.14159 rad | |||||
| θd | ±0.0001 deg - 180.000 deg | |||||
| Δ% | ±0.0001% - 999.999% | |||||
| Multi-function parameter list scanning | 1601 points, average can be set for each point, each point can be sorted individually
Scan parameters: measurement parameters, test frequency, AC voltage, AC current, DC BIAS voltage, DC BIAS current |
|||||
| Graphic Scanning | parameters | Frequency, ACV, ACI, DCV, DCI | ||||
| typology | Logarithmic, linear, frequency segmentation | |||||
| check numbers | 2-1601 | |||||
| channel number | 4 | |||||
| number of curves | 4 bars/channel | |||||
| split-screen display | 14 split-screen methods, 16 curves | |||||
| equivalent circuit analysis | 3-element models: 4, 4-element models: 3 | |||||
| Crystal Oscillator Analysis | there are | |||||
| Curve Trajectory Comparison | there are | |||||
| sorting | LCR mode with 10 levels of sorting; Scanning mode with separate sorting for each curve | |||||
| connector | RS232C, USB HOST, USB DEVICE, LAN, GPIB, HANDLER, VGA, HDMI | |||||
| Power-on warm-up time | 60 minutes. | |||||
| Input Voltage | 100-120VAC/198-242VAC selectable, 47-63Hz | |||||
| power wastage | Maximum 150VA | |||||
| Dimensions (WxHxD) mm3 | 428x220x325 | |||||
| weights | 14.5kg | |||||
■ Measurement frequency: 10Hz-50MHz
■ Basic impedance accuracy: 0.08%, typical value ≤ ±0.045%
■ High precision: using automatic balanced bridge technology, four-end pair test configuration
■ High stability and consistency
■ High speed: up to 2.5ms test speed
■ High resolution: 10.1-inch capacitive touch screen with 1280*800 resolution
■ Five test modes: point test, list scan, graphic scan, equivalent circuit, and crystal oscillator analysis
■ 1601-point multi-parameter list scanning function
■ Four-parameter measurements
■ Automatic Level Control (ALC) function
■ 4-channel graphic scanning function, each channel can display 4 curves, channels and curves are
14 split-screen displays
■ Analysis functions: equivalent circuit analysis, crystal oscillator analysis, piezoelectric analysis,
Curve Trajectory Comparison, Dielectric Constant Analysis (optional), Magnetic Permeability Analysis (optional)
■ Powerful sorting: 10-speed sorting in LCR mode
■ Graphical scanning mode with individual sorting for each curve, frequency segment scanning and sorting
■ High compatibility: Supports SCPI instruction set, compatible with KEYSIGHT E4990A,
e4980a, e4980al, hp4284a
■ Passive components:
Evaluation of impedance parameters and performance analysis of capacitors, inductors, cores, resistors, piezoelectric devices, transformers, chip assemblies, and network components.
■ Semiconductor components:
Test and analysis of parasitic parameters of LED driver integrated circuits; C-VDC characteristics of varactor diodes; analysis of parasitic parameters of transistors or integrated circuits
■ Other components:
Impedance evaluation of printed circuit boards, relays, switches, cables, batteries, etc.
■ Medium material:
Evaluation of dielectric constants and loss angles of plastics, ceramics and other materials
■ Magnetic materials:
Evaluation of permeability and loss angle of ferrites, amorphous and other magnetic materials
■ Semiconductor materials:
Dielectric constant, electrical conductivity and C-V properties of semiconductor materials
■ LCD unit:
C-V properties such as dielectric constant, elasticity constant, etc.
| standard equipment | |||||
| Accessory Name | model number | ||||
| fixture (machining) | TH26047 | ||||
| Four ends with card fixture | TH26005D | ||||
| 100Ω standard resistor | TH26082A | ||||
| short circuit board | TH26010 | ||||
| TH2851 Calibration Kit | TH2851 Calibration Kit | ||||
| optional | |||||
| Accessory Name | model number | ||||
| Magnetic Ring Clamp (Piece Type) | TH26007A | ||||
| Magnetic Ring Fixture | TH26008A | ||||
| Patch Test Leads with Boxes | TH26009B | ||||
| fixture (machining) | TH26048 | ||||
| Test Fixture | TH26062 | ||||
| Test Fixture | TH26063 | ||||
| High Frequency Patch Fixture | TH26108C | ||||
| Dielectric Test Fixture | TH26077 | ||||
| TH2851-7mm Fixture | TH26086 | ||||
| TH2851 Magnetic Permeability Test Fixture | TH26088 | ||||
| TH2851-7mm Fixture-Calibration Kit | TH26086-CAL | ||||
Comprehensive test and measurement service provider-Shenzhen Weike Electronic Technology Co.