■ modular system
The new 8460 is compatible with most applications. If your application software changes, the 8460 can be upgraded with a different test board (4 test boards available)
■ Multiple analysis functions
The new 8460 provides a wide range of automatic test data, multiple triggers, and power analysis modes, making it easy to analyze complex signals.
■ Easy-to-operate panoramic touch screen
With the 15.4-inch touchscreen, the experience of using the 8460 is like playing a game: the HMI is intuitively designed, all menus are simple and clear, and instructions for use can be displayed on the instrument.
▪ Instrument networking
With 6 USB ports, you can operate the data logger remotely or download the recorded data via LAN port or WIFI communication. Using the Virtual Network Computing software (not standard), you can browse and control the 8460 from your PC or tablet just as if you were using it face-to-face.
Common features (applicable to all such models) | |
demonstrate | |
15.4-inch TFT backlit touch screen | |
Resolution 1280*800 | |
Functions with XY coordinate display | |
Functions: zoom, cursor, inter-cursor zoom | |
Mathematics and proportional functions (Y=aX+B) | |
20 automatic measurements | |
save | |
Storage Settings | |
save | 128M words per segment |
internal hard disk | 500Gb, high speed transmission (6M/s) |
Interfaces and inputs/outputs | |
connector | Six USB (2 on front panel, 4 on rear panel) |
logical channel | 16 logic channels (max. voltage: 24V, input resistance 4.7kΩ) |
Sensor feed | 12V/0.2A max (non-current) |
Alarm output | 3 outputs, one relay (24V/100mA), two 2xTTL 5V |
Power analysis function (This function can be used with a universal board and accessories for current measurement.) | |
reticulation | Single-phase, two-phase, three-phase |
frequency | 50-60Hz, 400Hz, 1000Hz |
demonstrate | Oscilloscope Mode, Fresnel Mode |
harmonic (wave with frequency an integer multiple of the fundamental) | Calculation of up to 50th harmonic with logging function |
surveying | 24 test parameters: voltage, current (average, RMS, peak), peak factor, power (useful, useless, apparent), power factor, harmonics, THD, DF, frequency, energy ...... |
Routine and external environment | |
power supply | 90VAC - 264VAC, 47Hz - 63Hz |
power consumption | Maximum 230VA, except printer 60W |
operating temperature | 0°C - 40°C |
Storage temperature | -20°C - 60°C |
Maximum working humidity | Maximum 80% |
Dimensions (without extended enclosure) | 298*394*218mm |
Weight (one board) | 11kg (plus 10kg for extended housing) |
Specifications - Universal Input Board | ||
channel number | 6 channels per board | |
input voltage | ||
DC Voltage Range | 1mV - 1000V | |
Maximum Offset | Except ±5 1000V | |
accurate | ±0.1% ±10V ±0.2% Offset | |
TRMS AC+DC | 200mV - 500V | |
Bandwidth (-3dB) | 5Hz - 100kHz | |
crest factor | 2.2 | |
frequency | ||
(level of) sensitivity | 300mV rms min. | |
duty cycle | 10% | |
frequency range | 10Hz - 100kHz | |
Basic Precision | 0.2% Full Scale | |
Maximum Input Voltage | ±500V DC or 440V AC (sine) | |
temp | ||
transducers | operating temperature | realm |
J | -20°C - 1200°C | 20°C - 2000°C |
K | -250°C - 1370°C | 20°C - 2000°C |
T | -200°C - 400°C | 20°C - 500°C |
S | -50°C - 1760°C | 50°C - 2000°C |
B | -200°C - 1820°C | 50°C - 2000°C |
E | -250°C - 1000°C | 20°C - 1000°C |
N | -250°C - 1300°C | 20°C - 1000°C |
W5 | -0°C - 2320°C | 50°C - 2000°C |
accurate | Cold end compensation: ±1.25°C | |
sampling | ||
resolution (of a photo) | 14bit | |
Sampling speed | 1M/s per channel | |
Storage length | 32M words per paragraph, up to 128 paragraphs | |
trig | Positive Edge, Negative Edge, Logic Input, Delay, Go No Go | |
pre-trigger | -100% - 100% | |
bandwidths | ||
Analog Input Bandwidth | 1V: 100kHz
≤50mV : 50kHz min. |
|
Programmable Digital Filtering | 10Hz, 100Hz, 1kHz, 10kHz | |
Input Impedance (DC) | 25MΩ
≥1V: 1MΩ |
|
Input Capacitance | 150pF typ. | |
Maximum Input Voltage | Between one channel and enclosure ground: ±500V
Between two ports of a channel: ±500V |
|
Isolated enclosure ground and channel | >100MΩ at 500V DC |
Specifications - 6 Isolated High Voltage Channel Boards | |
channel number | 6 |
dc voltage | 100mV - 2000V |
Maximum Offset | ±5 range (limited to 2000 V maximum) |
accurate | ±0.2% ±0.2% offset |
Maximum RMS AC+DC Voltage | 1000V AC |
Bandwidth (-3dB) | 26kHz (range dependent) |
crest factor | 2.2 (up to 200 V peak)
|
Input Impedance | <10V: 11MΩ
≥10V: 10MΩ |
surety | CAT III - 1000V, CAT IV - 600V |
frequency | |
(level of) sensitivity | 100mVrms. |
duty cycle | 10% min. |
frequency range | 10Hz - 100kHz |
Basic precision | ±0.02% Full Scale |
sampling | |
resolution (of a photo) | 14bits |
Sampling speed | Maximum 1M/s per channel |
bandwidths | |
Analog Input Bandwidth | ≥100V: 26kHz
10V - 100V: 20kHz <10V: 3kHz |
Programmable Analog Filtering | 10kHz, 1kHz, 100Hz (slope 60dB/10 cells) |
Specifications - Multilayer Printed Boards | ||
channel number | 12 channels per board | |
input voltage | ||
DC Voltage Range | 1mV - 50V | |
Maximum Offset | ±5 | |
accurate | ±0.1% ±10μV ±0.1% offset | |
TRMS AC+DC | 200mV - 50V | |
Bandwidth (-3dB) | 5Hz - 100Hz | |
crest factor | 2.2 | |
temp | ||
transducers | operating environment | realm |
PT100(2,3,4Fils) | -200°C - 850°C | 20°C - 1000°C |
J | -20°C - 1200°C | 20°C - 2000°C |
K | -250°C - 1370°C | 20°C - 2000°C |
T | -200°C - 400°C | 20°C - 500°C |
S | -50°C - 1760°C | 50°C - 2000°C |
B | -200°C - 1820°C | 50°C - 2000°C |
E | -250°C - 1000°C | 20°C - 1000°C |
N | -250°C - 1300°C | 20°C - 1000°C |
WS | 0°C - 2320°C | 50°C - 2000°C |
Accuracy | Cold end compensation: ±1.25°C | |
sampling | ||
resolution (of a photo) | 16Bit | |
Sampling speed | Up to 200μs (5K samples/s) | |
Storage length | 32M words per paragraph, up to 128 paragraphs | |
trig | Positive Edge, Negative Edge, Logic Input, Delay, Go No Go | |
pre-trigger | -100% - 100% | |
bandwidths | ||
Analog input bandwidth (-3dB) | 1kHz at -3dB | |
Programmable Digital Filtering | 0.1Hz - 50Hz | |
Input Impedance (DC) | >5V: 2MΩ | |
Input Capacitance | ≤5V: 10MΩ (150pF) | |
Maximum Input Voltage | Between one channel and enclosure ground: ±50V
Between two ports of a channel: ±50V All inputs are differential, non-isolated |
|
Normal Mode Voltage (Max) | 5V: ±50V |
Specifications - Strain Gauge Plates | ||||
conduit | 6 channels (fully insulated) | |||
surveying | Strain gauge, voltage, thermocouple, and optional external shunt | |||
importation | Differential, Fully Isolated | |||
Input Impedance | <1V: 2MΩ
≥1V: 1MΩ |
|||
Maximum Input Voltage | 200V DC (between input and ground or between ground and enclosure) | |||
Input Voltage | ±50V | |||
incommunicado | >100MΩ at 500V (between channel and enclosure) | |||
Input Connector | Quick plug, 6 connectors per channel | |||
All accuracies are filtered at 1Hz | ||||
Voltage measurement | ||||
range | 50V | |||
bottom range | 1mV | |||
Maximum Offset | ±50V limited to ±5 range | |||
accurate | ±0.1% full scale, ±10μV, ±0.1% deviation | |||
resolution (of a photo) | 16bit | |||
drift out of tune | 100ppm/°C±1μV/°C | |||
Sampling speed | 100kHz (or 10μs) | |||
static (in a signal) | <30μV in unfiltered state | |||
Strain gauge measurements | ||||
This component is μSTR (microstrain) - 2000μSTR = 1mV/V | ||||
bridge (networking) | Full bridge (4 and 6 wires) Half bridge | |||
Automatic balancing range | ±25000μSTR | |||
Bridge Supply Voltage | 2V and 5V (symmetrical ±1V and plus or minus 2.5V) | |||
Meter rate | 2 (adjustable from 1.8 to 2.2) | |||
range | 50000μSTR | |||
minimum extent | 1000μSTR | |||
maximum deviation | ±50000μSTR | |||
accurate | ±0.1% full scale, ±5μSTR, ±0.1% deviation | |||
resolution (of a photo) | 16bit | |||
Sampling speed | 100kHz (or 10μs) | |||
drift out of tune | 100ppm/°C±1μV/°C | |||
bandwidths | ||||
3dB | >18kHz | |||
analog (device, as opposed digital)
filtering radio waves (i.e. pick out one frequency) |
Lowpass (60dB per 10 cells) | 1kHz, 100Hz, 10Hz | ||
Low Pass (Digital) | 1Hz, 0.1Hz, 0.01Hz, 0.001Hz | |||
temperature measurement | ||||
Cold end compensation for J, K, T, S, N, E, W5 thermocouples: ±1.25°C | ||||
transducers | Maximum possible reach | realm | ||
J | -210°C - 1200°C | 20°C - 2000°C | ||
K | -250°C - 1370°C | 20°C - 2000°C | ||
T | -200°C - 400°C | 20°C - 500°C | ||
S | -50°C - 1760°C | 50°C - 2000°C | ||
B | 200°C - 1820°C | 50°C - 2000°C | ||
E | -250°C - 1000°C | 20°C - 1000°C | ||
N | -250°C - 1300°C | 20°C - 1000°C | ||
W5 | 0°C - 2320°C | 50°C - 2000°C |
Test panels with options (plus * as factory option) | |
984405500 | 16 independent logical channel modules |
910007000 | logical channel connection |
984402000 | 12-Channel Multiplexer Board |
984401000 | 6 independent channel general purpose boards |
984402500 | 6 independent channels of strain gauges/temperature plates |
916006000 | 6 independent channel high voltage boards |
916003000 | IRIG board* |
916004500 | Wifi communication option |
current clamp | |
A1257 | Tool kit plus 3 bendable pliers 30A/300A/3000A AC for 3-phase testing |
A1287 | Bendable pliers 30A/300A/3000A AC |
SP201 | Current clamp 200A AC, 10mV/1A, D 15mm |
SP221 | Current clamp 10A AC, 100mV/1A, D 15mm |
SP230 | Current clamp 1200A AC, 10mV/1A, D 50mm |
SP261 | Current clamp 1200A AC+DC, 1mV/1A, D 50mm |
SP270 | Current clamp 2000A AC, 1mV/1A, D 70mm |
splitter | |
910007100 | Maximum shunt 0.01Ω 3A |
910007200 | Maximum shunt 0.1Ω 1A |
989006000 | Maximum shunt 1Ω 0.5A |
912008000 | Maximum shunt 10Ω 0.15A |
989007000 | Maximum shunt 50Ω 0.05A |
207030301 | Maximum shunt 0.01Ω 30A |
207030500 | Maximum shunt 0.001Ω 50A |
transport case | |
914007500 | For use with 8460 (without additional housing) |
FLEXPROanalysis software | |
100081 | Flexpro* View (Basic) |
100082 | Flexpro* View (full version) |
■ 6-36 analog channels, 16 logic channels
■ Maximum 1μs sampling speed per channel
■ General-purpose board bandwidth: 100kHz
High voltage board bandwidth: ≥100V: 26kHz
10V - 100V: 20kHz
<10V: 3kHz
Multiplexing board bandwidth: 1kHz
Strain gauge plate bandwidth: >18kHz
■ 500GB internal hard drive
■ 128M RAM
■ Test boards
-6 independent channels universal input 500V AC or 1000V DC
-12 channel multiplexer board (voltage, temperature, pt100)
-6 independent channels for strain gauges to test voltage, pt100 and thermocouples
-6 independent channels of 1000V AC* or 2000V DC*
■ 15.4-inch panoramic TFT touch screen
■ Interface: Ethernet, USB*6, VGA
■ Energy analysis of single and binary networks (50Hz, 60Hz, 400Hz, 1kHz)
■ IRIG board option
■ WiFi option
■ IEC1010: CAT III - 600V