● Wide range signal source from 1mHz to 200kHz for ion motion and solution resistance measurements.
Continuous measurement and high-speed inspection of LCR measurement and scanning measurement can be realized with one machine.
● Measurement of the internal resistance generated by the unloaded state of the battery is possible.
● High-speed measurement of up to 2ms, realizing high-speed scanning measurement.
:: Basic accuracy of ±0.05%, which can be applied from part inspection to R&D measurement.
:: Suitable for resistance (LCR) measurements of electrochemical parts and materials such as Cole-Cole diagrams, equivalent circuit analysis, etc.
Note: The main unit does not come with a test fixture. Please select the test fixtures and probes from the options.
※ The connection cable for RS-232C can be used with a normal cross cable that corresponds to the interlock. the RS-232C connection cable 9637 can be used only when hardware process control is not used.
※ The connection cable for RS-232C can be used with a normal cross cable that corresponds to the interlock. the RS-232C connection cable 9637 can be used only when hardware process control is not used.
Basic parameters
| measurement mode | LCR mode, continuous measurement (LCR/analysis), analysis mode (frequency or level sweep, temperature characterization, equivalent loop analysis) |
|---|---|
| measured parameter | Z, Y, θ, Rs (ESR), Rp, Rdc (DC resistance), X, G, B, Cs, Cp, Ls, Lp, D (tanδ), Q, T, σ (electrical conductivity), ε (dielectric constant) |
| measuring range | 100mΩ~100MΩ, 10 ranges (All parameters are specified by Z) |
| Display range | Z, Y, Rs, Rp, Rdc, X, G, B, Ls, Lp, Cs, Cp, σ, ε : ± (0.000000 [unit] to 9.999999G [unit]) Absolute values are displayed for Z and Y only. θ: ± (0.000° to 999.999°), D: ± (0.000000 to 9.999999) Q: ± (0.00 to 999999.99), Δ%: ± (0.0000% to 999.9999%) T: -10.0°C to 99.9°C |
| Basic precision | Z: ±0.05% rdg. θ: ±0.03° |
| Measurement frequency | 1 mHz to 200 kHz (1 mHz to 10 Hz step) |
| Measured signal level | Normal mode: V mode, CV mode: 5 mV to 5 V rms, 1 mV rms step CC mode: 10 μA to 50 mA rms, 10 μA rms steps Low impedance high precision mode: |
| output resistance | Normal Mode: 100 Ω, Low Resistance High Accuracy Mode: 25 Ω |
| demonstrate | Color TFT 5.7 inches, can be set to display ON/OFF |
| measuring time | 2ms (1kHz, FAST, display OFF, representative value) |
| functionality | DC Bias Measurement, DC Resistance Temperature Compensation (Standard Temperature Conversion Display), Temperature Measurement, Battery Measurement (Automatic DC Bias Addition Method), BIN Measurement (Sorting Function), Panel Read-Save, Memory Function |
| connector | EXT I/O (processor), USB communication, USB storage Option: One of RS-232C/GP-IB/LAN can be installed. |
| power supply | AC 100 to 240 V, 50/60 Hz, 50 VA max |
| Volume and weight | 330W × 119H × 168D mm, 3.1 kg |
| attachment (email) | Power cord × 1, Instruction manual × 1, CD-R (communication manual, operating software) × 1 |
Comprehensive test and measurement service provider-Shenzhen Weike Electronic Technology Co.