:: For short-time measurements in locations where there are no temperature variations
● Confirmation of measured values on site by means of a display unit
● Four types of outputs can be made to a memory recorder or data acquisition instrument depending on the application (used with a display unit) Waveform output, RMS output, peak output, and frequency output.
Basic parameters (accuracy guaranteed for 3 years, adjusted accuracy guaranteed for 3 years)
| Rated measuring current | AC/DC 600 A |
|---|---|
| Maximum Input Current | 600 A (with frequency derating) |
| maximum peak value | 900 A peak |
| frequency bandwidth | DC~10 kHz (-3dB) |
| Phase Representation Accuracy | ±1.8 deg. (DC < f ≤ 66 Hz) |
| output rate | 1 mV/A |
| Maximum rated voltage to ground | AC/DC 600 V (CAT IV) AC/DC 1000 V (CAT III) |
| Measurable conductor diameter | φ 33 mm or less |
| output connector | HIOKI PL14 |
| Temperature range | -25℃~65℃ |
| Dust and water resistance* | Jaws, barrier: IP50 / Hand-held part: IP54 (only when measuring insulated conductors, do not measure in wet conditions) |
| Volume and weight | 64W × 160H × 34D mm, 320 g, cord length 2.5 m |
| attachment (email) | not have |
*Waterproofing enables measurements to be performed, but the risk of electric shock increases when measuring electrically charged parts when the unit is wet.
Comprehensive test and measurement service provider-Shenzhen Weike Electronic Technology Co.