The selection criteria for a VNA depends heavily on the application or usage scenario. Typically, selection criteria include frequency range, number of ports required (from 2 to 24 (multiport) and even 48 (using a switching matrix)), output power, and instrument class from economy to handheld to high-end.
Key Specifications of Vector Network Analyzers
Vector network analyzers are both signal generators and receivers, so they have a large number of very necessary technical specifications. In this section, you will learn about some of the key technical specifications of network analyzers.
Maximum frequency
The maximum frequency of a VNA is the highest frequency it can measure. The receiver side of a network analyzer has an analog-to-digital converter (ADC) that converts the input signals into digital format. These signals can then be analyzed and displayed. However, the ADC does not have the ability to convert signals in the RF range, so the incident signal must be downconverted to its operating frequency. This operating frequency is called the intermediate frequency (IF).
dynamic range
Dynamic range is the range of power over which the response of a component can be measured.
This figure shows two different ways of defining dynamic range. The system dynamic range is the value used for instrument specifications.
The system dynamic range is the function of the instrument when no boost amplifier is used and the gain of the device under test is not taken into account. The maximum source power of an instrument is its maximum power level, Pref
The receiver dynamic range is the dynamic range of the instrument when power amplification is used. Unlike using the source power as the maximum power level, this specification is based on the maximum power Pmax that can be measured at the receiving end of the instrument.
Defining the dynamic range
A trace measured by the bandpass filter S21, which shows the dynamic range of the instrument, is shown in the lower left-hand plot. The upper limit of the trace is relatively flat and the lower limit contains some noise. Let's look at what factors determine the shape of these boundaries.
The maximum power level of the dynamic range is determined by the upper limit of the source power level and the compression point of the receiver.
The mixers and amplifiers that make up a receiver can only handle so much power before they reach saturation, or before they reach maximum output. When a device is in the saturation region, there is no longer a linear relationship between its input and output.
The saturation of the amplifier can be seen in the right-hand diagram below. At input powers higher than 1W, the actual output (red) will deviate from the desired output (green). This phenomenon is called compression. The receiver cannot capture the signal from any device output above its compression point. This limitation of the input power constitutes the upper limit of the dynamic range.
Dynamic range of traces In the gain compression plot, the ideal linear transfer function of the amplifier is shown in green and the true transfer function is shown in red.
output power
Output power reflects how much power the VNA's signal generator and tester can transmit into the device under test. It is expressed in dBm and is referenced to 50Ω impedance to match the characteristic impedance of most RF transmission lines.
High output power is useful for improving the signal-to-noise ratio of a measurement or determining the compression limits of the device under test.
Many active devices, such as amplifiers, require extremely challenging linear and nonlinear high power measurements beyond the power limits of the network analyzer.
trace noise
Trace noise is the superimposed noise you see forming on the response of the device under test due to random noise in the system. It can make the signal look less smooth and even a little jittery.
Trace noise can be eliminated by increasing the test power, decreasing the bandwidth of the receiver, or averaging.
An oscilloscope uses a probe to connect to a test point in a circuit and amplify or attenuate the voltage at that point. The subsequent signal path depends on the oscilloscope, but in general, analog circuits condition the signal and analog-to-digital converters digitize it for further evaluation, processing, and analysis. ,
Keysight PathWave BenchVue software includes and hosts a variety of applications. BenchVue applications help you control your test instruments and streamline your testing process. PathWave BenchVue software allows you to connect to, control, and capture data from a growing range of Keysight instruments. Most function generators come with PC application software that allows you to connect...
Want to know how to use the current probe? Please first understand the following basic usage of the current probe and precautions for use. Since electricity is invisible, it is not possible to immediately determine whether a malfunction has occurred. In some cases, current measurements are required on a daily basis to prevent malfunctions, as well as to identify the cause of malfunctions when they do occur. Digital multimeters, clamp meters, and current probes can be used...
Why do you need to measure current? In this article, we will cover the reasons for measuring current, how to measure current and what to look for when measuring current. We cannot see the current that is flowing with our own eyes. Therefore, specialized measuring instruments are needed to measure current and other parameters. But why do we need to measure current? And how should we measure it? This article will detail the reasons for the need to measure current...
Hello!sign in