TH1991/TH1992 series precision source/measurement power supply can output and measure voltage and current at the same time. Current source, voltage source, voltmeter and ammeter functions are integrated in the instrument, and each function can be switched at will.

The TH1991/TH1992 series of precision source/measurement units can output up to ±210V DC voltage, ±3A DC current, and ±10.5A pulsed current, with a minimum of 10fA/100nV power supply and measurement resolution, and support high-speed sampling to generate arbitrary waveforms.

TH1991/TH1992 series precision source/measurement unit series adopts 7-inch capacitive touch screen, with Linux operating system as the underlying, interactive graphical user interface and various display modes, and built-in diodes, transistors, MOS tubes, and IGBT devices such as I/V curve scanning function, without the need to connect to the upper computer can be completed without the IV function of the test, which can significantly improve the test efficiency. It can significantly improve the testing efficiency.

The TH1991/TH1992 series of precision source/measurement units' comprehensive, integrated power supply and measurement functions are ideal for testing semiconductors, active/passive devices, and a variety of other devices and materials.

The TH1991/TH1992 series of precision source/measurement units are used in a wide range of applications in R&D and educational applications, industrial development, test and manufacturing.

Categories. , Tags. , , Brand.

Inquire about this product

TH1991/TH1992 series precision source/measurement power supply can output and measure voltage and current at the same time. Current source, voltage source, voltmeter and ammeter functions are integrated in the instrument, and each function can be switched at will.

The TH1991/TH1992 series of precision source/measurement units can output up to ±210V DC voltage, ±3A DC current, and ±10.5A pulsed current, with a minimum of 10fA/100nV power supply and measurement resolution, and support high-speed sampling to generate arbitrary waveforms.

TH1991/TH1992 series precision source/measurement unit series adopts 7-inch capacitive touch screen, with Linux operating system as the underlying, interactive graphical user interface and various display modes, and built-in diodes, transistors, MOS tubes, and IGBT devices such as I/V curve scanning function, without the need to connect to the upper computer can be completed without the IV function of the test, which can significantly improve the test efficiency. It can significantly improve the testing efficiency.

The TH1991/TH1992 series of precision source/measurement units' comprehensive, integrated power supply and measurement functions are ideal for testing semiconductors, active/passive devices, and a variety of other devices and materials.

The TH1991/TH1992 series of precision source/measurement units are used in a wide range of applications in R&D and educational applications, industrial development, test and manufacturing.

1676077247259395-6

Functional Features

A. Four-quadrant output

TH199X series available in four quadrants, maximum output range (some models only)

1676077301298812-3

B. Multiple ways of working

The TH199X can operate in the following operating modes

1676077325136354-5

C. Static Characterization of Semiconductor Components

TH199X series precision source/measurement unit without the need for host computer software, can be set up directly in the instrument interface and generate semiconductor components such as diodes, transistors, MOS tubes, IGBTs, etc. IV characteristic curve, and you can directly call the commonly used device library, quickly complete the test.

1) Diode IV Characterization Test

1676077368130997-3

1676077379357680-6

2) Field effect tube output characteristic curve

1676077397261202-4

1676077408803747-2

3) Triode input-output characteristic curve

1676077430122403-3

1676077441838775-5

Product Model TH1991C TH1991B TH1991A TH1991 TH1992B TH1992A TH1992
demonstrate
monitor (computer)                                     7-inch capacitive touch color LCD monitor with 800 x 480 resolution
Key indicators
channel number 1 2
power (output) 31.8W
maximum output input voltage ±63V ±210V
amps direct current (D.C.) ±1.515A ±3.03A
impulse ------ ------ ±10.5A ±10.5A ------ ±10.5A ±10.5A
power supply highest degree figure 5 1/2 6 1/2 5 1/2 6 1/2
minimum resolution input voltage 1μV 100nV 1μV 100nV
amps 1pA           100fA                1pA 10fA 100fA         1pA 10fA
surveying highest degree figure 6 1/2
minimum resolution input voltage 100nV
amps 100fA 10fA 100fA 10fA 10fA 100fA 10fA
Voltage range 200mV

-60V

200mV-200V
Minimum Interval 50μs 20μs          10μs         1μs           20μs          10μs          1μs
Voltage measurement/voltage source (accuracy: % of reading + bias)
range (of scales or measuring equipment) ±200mV Measurement resolution 100nV
accurate ±(0.015% + 225μV)
±2V Measurement resolution 1μV
accurate ±(0.02% + 350μV)
±20V Measurement resolution 10μV
accurate ±(0.015% + 5mV)
±200V Measurement resolution 100μV
accurate ± (0.015% + 50mV)
Current Measurement/Current Source
range (of scales or measuring equipment) ±10 nA Measurement resolution 10fA (TH1991A/TH1992A do not have this range)
accurate ± (0.10 % + 50 pA)
±100nA Measurement resolution 100fA
accurate ±(0.06% + 100pA)
±1μA Measurement resolution 1pA
accurate ± (0.025% + 500pA)
±10μA Measurement resolution 10pA
accurate ±(0.025% + 1.5nA)
±100μA Measurement resolution 100pA
accurate ±(0.02% + 25nA)
±1mA Measurement resolution 1nA
accurate ±(0.02% + 200nA)
±10mA Measurement resolution 10nA
accurate ±(0.02% + 2.5μA)
±100mA Measurement resolution 100nA
accurate ±(0.02% + 20μA)
±1A Measurement resolution 1μA
accurate ±(0.03% + 1.5mA)
±1.5A Measurement resolution 1μA
accurate ±(0.05% + 3.5mA)
±3A Measurement resolution 10μA
accurate ±(0.4% + 7mA)
±10A Measurement resolution 10μA
accurate ±(0.4% + 25mA)
Pulse source (pulse width is the time from the rising edge of 10% to the falling edge of 90%, base level: pulse low level, peak level: pulse high level)
Minimum programmable pulse width 50μs
Pulse Width Programming Resolution 1μs
DC or pulse maximum voltage 210V Maximum peak current 0.105A
Maximum base current 0.105A
pulse width 50μs - 99999.9s
Maximum Duty Cycle 99.9999%
21V Maximum peak current 1.515A
Maximum base current 1.515A
pulse width 50μs - 99999.9s
Maximum Duty Cycle 99.9999%
6V Maximum peak current 3.03A
Maximum base current 3.03A
pulse width 50μs - 99999.9s
Maximum Duty Cycle 99.9999%
pulse only 200V Maximum peak current 1.515A
Maximum base current 50mA
pulse width 50μs - 2.5ms
Maximum Duty Cycle 2.5%
180V Maximum peak current 1.05A
Maximum base current 50mA
pulse width 50μs - 10ms
Maximum Duty Cycle 2.5%
6V Maximum peak current 10.5A
Maximum base current 0.5A
pulse width 50μs - 1ms
Maximum Duty Cycle 2.5%
Resistance measurement (automatic resistance measurement mode, 4-wire, 2 V range)
range (of scales or measuring equipment) resolution (of a photo) 1μΩ
Test Current 1 A
Current range 1 A
total error 0.2% + 0.00035Ω
20Ω resolution (of a photo) 10μΩ
Test Current 100mA
Current range 100mA
total error 0.06% + 0.0035Ω
200Ω resolution (of a photo) 100μΩ
Test Current 10mA
Current range 10mA
total error 0.065% + 0.035Ω
2kΩ resolution (of a photo) 1mΩ
Test Current 1mA
Current range 1mA
total error 0.06% + 0.35Ω
20kΩ resolution (of a photo) 10mΩ
Test Current 100μA
Current range 100μA
total error 0.065% + 3.5Ω
200kΩ resolution (of a photo) 100mΩ
Test Current 10μA
Current range 10μA
total error 0.06% + 35Ω
2MΩ resolution (of a photo)
Test Current 1μA
Current range 1μA
total error 0.095% + 350Ω
20MΩ resolution (of a photo) 10Ω
Test Current 100nA
Current range 100nA
total error 0.18% + 3.5kΩ
200MΩ resolution (of a photo) 10Ω
Test Current 10nA
Current range 10nA
total error 1.08% + 35kΩ
connector RS232C, USB HOST, USB DEVICE, LAN, HANDLER
Environment and temperature
Operating temperature and humidity range 23°C±5°C
Storage temperature and humidity range 23°C±5°C
Precision guaranteed temperature and humidity 23°C±5°C
preheating time 60 minutes.
Change in ambient temperature 30% to 80%RH
calibration period first year
General indicators
power supply 90 V to 264 V, 47 Hz to 63 Hz, 250 VA maximum
power (output) 31.8W
Shelf Size 215mmx132mmx490mm
Overall dimensions 235mmx154mmx530mm
weights Approx. 8.5kg (single channel) / 10kg (dual channel)

■ Number of channels: 1

■ 高达±210V直流电压、±3A直流电流/±10.5A脉冲

■ 7-inch capacitive touchscreen with 800 x 480 resolution

▪ Linux operating system

■ Four-quadrant precision power output

■ Five basic modes: voltage source, current source, voltmeter, ammeter or resistance meter

Supports DC, pulse, scanning and list outputs.

■ Diodes, transistors, MOS tubes and IGBTs to quickly generate the characteristic curve of commonly used devices, time domain waveform scrolling display function

■ Combined 2-wire measurement/ 4-wire measurement

■ Output filter time constant (or cutoff frequency) can be freely set to realize arbitrary frequency response output.

■ Math operation function, sliding average filter function, deviation deduction function

■ Delta low-resistance test method, effectively compensating for measurement errors caused by thermoelectric potentials

■ 14-speed sorting function with Grading and Sorting modes.

■ TH199X series up to 1,000,000 points/second sampling rate, minimum sampling interval 1μs

 

■ Pulse width of TH199X series pulse outputs down to 50μs

■ TH199X Series Arbitrary Waveform Generation and List Scan Function (Minimum 1μs Interval)

■ Standard with the host computer software, easy to control and data acquisition on the computer

■ Semiconductor, Discrete and Passive Component Testing

Diodes, laser diodes, LEDs

Photodetectors, sensors

Field Effect Tubes, Triodes

ICs (ICs, RFICs, MMICs)

Resistors, Varistors, Thermal Electronics, Switches

■ Precision electronics and green energy device testing

photovoltaic cell

power semiconductor

batteries

motor vehicles

Medical Instruments

Power and DC Bias Sources for Board Level Testing

▪ Research and education

New Materials Research

Nanodevice Characterization

giant magnetoresistance (GMR)

Organic equipment

Any precision I/V source or measurement

standard equipment
Accessory Name model number
Test cable on both ends TH26050B
Banana Head to RF (Both Ends) Test Kit TH26087B
Four-terminal test cable (five-wire) TH26050C

 

optional
Accessory Name model number
Banana Head to RF (Quad) Test Kit TH26087D
Banana Head to Quad Coaxial Adapter (4-wire) TH26087C
shielding case TH26087E
Triaxial Cable TH26087F
Triaxial Cable TH26087G
Expand More
TH1991 High Accuracy Source MeterTH1991 High Accuracy Source Meter

Click to enquire

Product consultation and purchase:18218718851

I already understand