  • Maximum Output ±210V/ ±1.05A/ 22W
  • 4.3" TFT LCD, numeric keypad
  • quadrant work
  • I/V curve, I/t curve, V/t curve
  • Voltage source, current source, digital meter function
  • 0.012% Basic Measurement Accuracy, 6½ Digital Meter Test, 1μV/10pA/10μΩ Resolution
  • Built-in 4 sequence output modes (Stair, Log, SRC-MEM, Custom), up to 2500 points
  • SDM (Source Delay Measure)
  • 2/4/6 Wire Resistance Testing Remote V-source and Measure Sensing
  • Built-in Pass/Fail function
  • Interface: RS-232C, USBTMC, LAN, GPIB (optional)

Categories. , Tags. , Brand.

Inquire about this product

The GSM-20H10 provides ±210V / ±1.05A / 22W four-quadrant operation. In the first and third quadrants, they operate as power supplies to power the load. In the second and fourth quadrants, they operate as loads, dissipating power internally. Voltage, current and resistance values can be measured during power or load operation with an accuracy of 0.012% and a resolution of 1µV/10pA/10µΩ.

The GSM-20H10 supports a sampling speed of up to 50k points per second, which allows for more accurate analysis of the properties of the object to be measured. The large 4.3-inch screen displays all measurement settings, parameters and results. Support SDM (Source Delay Measure) function, which can be set to delay sampling when the signal changes, avoiding the signal before stabilization to be captured in and cause misjudgment. Built-in four program output modes (Linear, Log, SRC-MEM, Custom) can support up to 2500 points of program change output.

In terms of device protection, GSM-20H10 provides OVP/OTP protection modes; in the design of OVP, users can customize the range of OVP, while the OTP protection can effectively prevent errors caused by temperature offset during the testing process. In the interface section, this product supports standard SCPI commands and provides RS-232 , USBTMC , LAN, GPIB (optional) and other interfaces to meet the different interface needs of users.

Product Specification

Maximum Range Voltage ±210V
Current ±1.05A
Power 22W
Voltage Resolution 1µV
Current Resolution 10pA
Source DC Voltage Output Voltage ±21V/ ±1.05A, ±210V/±105 mA
Current Limit Min. 0.1% of range
Programming Resolution 1µV, ±200.000mV range
10µV, ±2.00000V range
100µV, ±20.0000V range
1mV, ±200.000V range
Programming Accuracy 1 ±(0.02%+600µV), ±200.000mV range
±(0.02%+600µV), ±2.00000V range
±(0.02%+2.4mV), ±20.0000V range
±(0.02%+24mV), ±200.000V range
Load Regulation 0.01% of range + 100µV
Line Regulation 0.01% of range
Overshoot <0.1% typical (full scale step, resistive load,10mA range)
Recovery Time (1000%Load Change) <250µs (within 0.1% plus load regulation errors, 1A and 100mA compliance.)
Ripple and Noise 4mV rms(20Hz~ 1MHz) 10mVpp(20Hz~ 1MHz)
Temperature Coefficient
(0°-18°C & 28°-50°C)
±(0.15 × accuracy specification)/°C
DC Current Output Current ±1.05A / ±21V, ±105 mA / ±210V
Voltage Limit Min. 0.1% of range
Programmed Source Resolution 10pA, ±1.00000µA range
100pA, ±10.0000µA range
1nA, ±100.000µA range
10nA, ±1.00000mA range
100nA, ±10.00000mA range
1µA, ±100.000mA range
10µA, ±1.00000A range
Programmed Source Accuracy 1 ±(0.035%+600pA), ±1.00000µA range
±(0.033%+2nA), ±10.0000µA range
±(0.031%+20nA), ±100.000µA range
±(0.034%+200nA), ±1.00000mA range
±(0.045%+2µA), ±10.00000mA range
±(0.066%+20µA), ±100.000mA range
±(0.27%+900µA), ±1.00000A range
Load Regulation 0 0.01% of range + 100pA
Line Regulation 0.01% of range
Overshoot <0.1% typical (1mA step, RL = 10kΩ, 20V range).
Temperature Coefficient
(0°-18°C & 28°-50°C)
±(0.15 × accuracy specification)/°C
General Output Settling Time2 100µs typical Time
Output Rise Time (±30%) 300µs, 200V range, 100mA compliance.
150V/µs, 20V range, 100mA compliance.
DC Floating Voltage Output can be floated up to ±250VDC
Remote Sense Up to 1V drop per load lead.
Compliance Accuracy Add 0.3% of range and ±0.02% of reading to base specification.
Range Change Overshoot 3 Adjacent range changes between 200mV, 2V and 20V ranges, 100mV typical.
Minimum Compliance Value 0.1% of range
Command Processing Time 4 Autorange On:10ms.Autorange Off:7ms.
Measurement  Voltage Input Resistance >10 GΩ
Measurement Resolution 1µV, ±200.000mV range
10µV, ±2.00000V range
100µV, ±20.0000V range
1mV, ±200.000V range
Measurement Accuracy ±(0.012%+300µV), ±200.000mV range
±(0.012%+300µV), ±2.00000V range
±(0.015%+1.5mV), ±20.0000V range
±(0.015%+10mV), ±200.000V range
Temperature Coefficient
(0°-18°C & 28°-50°C)
±(0.15 × accuracy specification)/°C
Current Voltage Burden (4-wire mode) < 1mV
Programmed Source Resolution 10pA, ±1.00000µA range
100pA, ±10.0000µA range
1nA, ±100.000µA range
10nA, ±1.00000mA range
100nA, ±10.00000mA range
1µA, ±100.000mA range
10µA, ±1.00000A range
Programmed Source Accuracy1 ±(0.029%+300pA)), ±1.00000µA range
±(0.027%+700pA), ±10.0000µA range
±(0.025%+6nA), ±100.000uA range
±(0.027%+60nA), ±1.00000mA range
±(0.035%+600nA), ±10.00000mA range
±(0.055%+6µA), ±100.000mA range
±(0.22%+570µA), ±1.00000A range
Temperature Coefficient
(0°-18°C & 28°-50°C)
±(0.1 × accuracy specification) / °C
Resistance Range Resolution Test current Accuracy
<2.00000Ω --- --- Source IACC+Meas.VACC
2.00000Ω 10µΩ --- Source IACC+Meas.VACC
20.0000Ω 100µΩ 100mA ±(0.1%+0.003 Ω), Normal
±(0.07%+0.001 Ω), Enhanced
200.000Ω 1mΩ 10mA ±(0.08%+0.03 Ω), Normal
±(0.05%+0.01 Ω), Enhanced
2.00000kΩ 10mΩ 1mA ±(0.07%+0.3 Ω), Normal
±(0.05%+0.1 Ω), Enhanced
20.0000kΩ 100mΩ 100µA ±(0.06%+3 Ω), Normal
±(0.04%+1 Ω), Enhanced
200.000kΩ 10µA ±(0.07%+30 Ω), Normal
±(0.05%+10 Ω), Enhanced
2.00000MΩ 10Ω 5µA ±(0.11%+300 Ω), Normal
±(0.05%+100 Ω), Enhanced
20.0000MΩ 100Ω 0.5µA ±(0.11%+1k Ω), Normal
±(0.05%+500 Ω), Enhanced
200.000MΩ 1kΩ 100nA ±(0.66%+10k Ω), Normal
±(0.35%+5k Ω), Enhanced
>200.000M Ω --- --- Source IACC+Meas.VACC
Temperature Coefficient
(0°-18°C & 28°-50°C)
±(0.15 × accuracy specification)/°C
Source I mode, Manual OHMS Total uncertainty = I source accuracy + V measure accuracy (4-wire remote sense).
Source V mode, Manual OHMS Total uncertainty = V source accuracy + I measure accuracy (4-wire remote sense).
6-wire OHMS mode Available using active ohms guard and guard sense. Max. guard Output Current: 50mA (except 1A range). Accuracy is load dependent.
Guard Output Impedance <0.1Ω in ohms mode
System Speed Maximum Range Change Rate 75/second
Maximum Measure Auto Range Time 40ms (fixed source) 6
Sequence reading rates7 (rdg./second) for 60Hz (50Hz) Speed NPLC/ Trig Origin Measure Source-Measure 9
TO MEM. TO GPIB TO MEM.        TO GPIB
Fast 0.01 / internal 2081 (2030) 1198 (1210) 1551 (1515) 1000 (900)
488.2 0.01 / external 1239 (1200) 1079 (1050) 1018 (990) 916 (835)
Medium 0.1 / internal 510 (433) 509 (433) 470 (405) 470 (410)
488.2 0.1 / external 438 (380) 438 (380) 409 (360) 409 (365)
Normal 1 / internal 59 (49) 59 (49) 58 (48) 58 (48)
488.2 1 / external 57 (48) 57 (48) 57 (48) 57 (47)
Speed NPLC/ Trig Origin Source-Measure Pass/Fail test 8,9 Measure Memory 9
TO MEM. TO GPIB TO MEM.        TO GPIB
Fast 0.01 / internal 902 (900) 809 (840) 165 (162) 164 (162)
488.2 0.01 / external 830 (830) 756 (780) 163 (160) 162 (160)
Medium 0.1 / internal 389 (343) 388 (343) 133 (126) 132 (126)
488.2 0.1 / external 374 (333) 374 (333) 131 (125) 131 (125)
Normal 1 / internal 56 (47) 56 (47) 44 (38) 44 (38)
488.2 1 / external 56 (47) 56 (47) 44 (38) 44 (38)
Single Reading Operation Rates (rdg./second) for 60Hz (50Hz) Speed NPLC/ Trig Origin Measure Source-Measure 9 Source-Measure Pass/Fail test 8,9
TO GPIB TO GPIB TO GPIB
Fast (488.2) 0.01 / internal 256 (256) 79 (83) 79 (83)
Medium (488.2) 0.1 / internal 167 (166) 72 (70) 69 (70)
Normal (488.2) 1 / internal 49 (42) 34 (31) 35 (30)
Component Interface Handler Time for 60Hz (50Hz): 8, 10 Speed NPLC/ Trig Origin Measure Source Pass/Fail test Source-Measure Pass/Fail test 9,11
TO GPIB TO GPIB TO GPIB
Fast 0.01 / internal 1.04 ms (1.08 ms) 0.5 ms (0.5 ms) 4.82 ms (5.3 ms)
Medium 0.1 / internal 2.55 ms (2.9 ms) 0.5 ms (0.5 ms) 6.27 ms (7.1 ms)
Normal 1 / internal 17.53 ms (20.9 ms) 0.5 ms (0.5 ms) 21.31 ms (25.0 ms)
System General Load Impedance Stable into 20,000pF typical
Differential mode Voltage 250 VPk
Common mode Voltage 250 VDC
Common mode Isolation >10GΩ, <1000pF
Over Range 105% of range, source and measure.
Max. voltage drop 5V
Max. Sense lead Resistance 1MΩ
Sense input Impedance >100G Ω
Guard offset Voltage <150µV, typical
Source output modes Fixed DC level, Memory List (mixed function), Stair (linear and log)
Source memory list 100 points max.
Memory buffer 5,000 readings @ 5 digits (two 2,500 point buffers). Includes selected measured value(s) and time stamp. Lithium battery backup(3 yr+ battery life).
Programmability IEEE-488.2 (SCPI)
5 user-definable power-up states plus factory default and *RST.
Digital I/O Connector Active low input.
Start of test, end of test, 3 category bits.
+5V@ 300mA supply.
1 trigger input, 4 TTL/Relay Drive outputs (33V @ 500mA, diode)
Remote Interface USB/GPIB/LAN/RS-232
Insulation Chassis and Terminal 20MΩ or above (DC 500V)
Chassis and AC cord 30MΩ or above (DC 500V)
Operation Environment Indoor use, Altitude: ≤ 2000m Ambient temperature: 0 ~ 40°C Relative humidity: ≤ 80%
Installation category: II, Pollution degree: 2
Storage Environment Temperature: -20˚C ~ 70˚C
Humidity: < 80%
Input Power 100-240VAC, 50-60Hz
Power Consumption 80W
Accessories CD User manual x1, Quick Start manual x1
Test lead GTL-207A x 1, GTL-108A x 1, ALLIGATOR CLIP x 2
Dimensions 214 (W) x 86 (H) x 356.5 (D) mm
Weight Approx. 4.8kg

Remarks

  1. Speed = Normal (1 PLC). For 0.1 PLC, add 0.005% of range to offset specifications, except 200mV, 1A ranges, add 0.05%. For 0.01 PLC, add 0.05% of range to offset specifications, except 200mV, 1A ranges, add 0.5%. For 0.01 PLC, add 0.05% of range to offset specifications, except 200mV, 1A ranges, add 0.5%.
  2. Required to reach 0.1% of final value after Command is processed. Resistive load. 10µA to 100mA range.
  3. Overshoot into a fully resistive 100kΩ load, 10Hz to 1MHz BW, adjacent ranges: 100mV typical, except 20V/200V.
  4. Maximum time required for the output to begin to change following the receipt of :SOURce:VOLTage|CURRent Command
  5. Reading rates applicable for voltage or current measurements, autorange off, filter off, display off, trigger delay = 0, and binary reading forma.
  6. Purely resistive lead. 1µA and 10µA ranges <65ms.
  7. 1000 point sweep was characterized with the source on a fixed rang.
  8. Pass/Fail test performed using one high limit and one low math limit.
  9. Includes time to re-program source to a new level before making measurement.
  10. Time from falling edge of START OF TEST signal to falling edge of END OF TEST signal.
  11. Command processing time of :SOURce:VOLTage|CURRent: TRIGgered Command not included.

standard equipment

User's Manual CD*1

Power cord*1

Test Lead GTL-207A*1

GTL-203A*1

GTL-204A*1

optional

SM-03. GPIB Card

Optional Accessories

SM-01 Digital I/O Adapter, Convert DB15 to DB9 + 8-pin micro-DIN
SM-02 Digital I/O Adapter, Convert DB15 to DB37 + 8-pin micro-DIN
GTL-246 USB Cable (USB 2.0 A-B Type, approx. 1200mm)
GTL-258 GPIB Cable (25 pin Micro-D Connector)

 

Ordering Information

GSM-20H10(with GPIB) High Accuracy Source Meter

GSM-20H10 High Accuracy Source Meter

Expand More
GSM-20H10 Digital Source MeterGSM-20H10 Digital Source Meter

Click to enquire

Product consultation and purchase:18218718851

I already understand